4.1.2
Considerations
The neighboring points of a line parallel to the viewing-point axis passing through the reference point represent points of equal size but of different luminances. Focusing on these points from the perspective of the effect of size on the depth perception of a small-field light source, a larger light source is perceived as being in front, conversely, a smaller light source is perceived as being behind. This result is consistent with the feature of retinal image size, the larger the image on the retina, the more it is perceived to be in the foreground. The neighboring points of a line parallel to L∗ axis passing through the reference point represent points of equal size but of different luminances. Analyzing these points to examine the effect of brightness on depth perception of small-field light sources, it can be inferred that points with brighter light are perceived as being in front, whereas darker ones are perceived as being behind. This result is consistent with the feature of light attenuation, when light at a greater distance appears darker.
Figure 9.
Next, we consider the relationship between the size and luminance in the depth perception of small-field light sources.
As the binary classification line is curved, the predominance of retinal image size or light attenuation could not be ascertained. However, several studies report that the total amount of light reflected from the eye influenced the selection.
We analyzed the relationship between retinal illuminance, a photometric quantity representing illuminance on the retinal surface, and the experimental results. The magnitude of retinal illuminance determined depth perception. When viewing a surface with luminance
L (cd/m
2) through a pupil of radius
r (mm), the retinal illuminance
Er(
td) can be expressed as:
We determined the retinal illuminance of a small-field light source with luminance
L (cd/m
2) and radius
a (mm). Initially, as the luminance of a small-field light source differs from that of an extended source,
L is not directly substituted into Eq. (
1) and the average of the surrounding luminance
L′ is computed. The area of the extended source projected onto the retina was considered as a circle with radius
A (mm). A schematic of this process is shown in Figure
9. Eq. (
2) expresses the luminous intensity I within a circle:
Figure 10.
Identical luminous intensity function. (a) Observations pertaining to reference point 1. (b) Observations pertaining to reference point 2. (c) Observations pertaining to reference point 3.
Thus, the average
Lave of the surrounding luminance of a small-field light source can be calculated as follows:
We consider a reference point with luminance
L0 (cd/m
2) and radius
a0 (mm) observed through a pupil diameter
r0 (mm). Using Eq. (
3), the luminance
L and radius
a of a small-field light source with a retinal illuminance equal to
Er0(
td) can be calculated.
Experimental results show that under 1
lx illumination, the pupil diameter ranged between 4 mm and 9 mm [
2], and under illumination of 0 (cd/m
2), it ranged between 4 mm and 6 mm [
23]. As it is impossible to conduct the experiment and measure the pupil diameter simultaneously and since values vary significantly from person to person, this study assumes that the pupil diameter
r varies with the luminous intensity of small-field light source. When the luminance and radius of the reference point are
L0 (cd/m
2) and
a0 (mm), respectively, the luminous intensity
I0 of the reference point is calculated as follows:
The function of radius
a and luminance
L of a point that has the same luminous intensity as this point is as follows:
Figure
10 illustrates the function of
L∗ and the viewing angle for points with same luminous intensity as the reference point determined using Eq. (
6) (purple in Fig.
8). Fig.
10 indicates that the luminous intensity is greater than that of the reference point for a larger viewing angle, and smaller for a smaller viewing angle. Fig.
10(a) corresponds to reference point 1; Fig.
10(c) corresponds to point 3. The values along the binary classification line obtained using SVM show that the luminous intensity is lower than the reference point for lower
L∗ values and higher for higher
L∗ values. Fig.
10(b) corresponds to reference point 2, and the values along the binary classification line obtained using SVM similarly indicate that the luminous intensity decreases as
L∗ decreases. However, at higher
L∗ values, the luminous intensity is almost equal to that of the reference point.
Figure 11.
Identical retinal illuminance function. (a) Observations pertaining to reference point 1. (b) Observations pertaining to reference point 2. (c) Observations pertaining to reference point 3.
It can be inferred from these results that
L∗ determines the difference in luminous intensity between the threshold and reference point. Moreover, assuming that the pupil diameter changes with the luminous intensity of small-field light sources, we can conclude that a lower
L∗ results in a larger pupil diameter, owing to lower luminous intensity, and a higher
L∗ results in a smaller pupil diameter, due to higher luminous intensity. Reiterating the inability to measure pupil diameter simultaneously with the experiment, this study assumed a linear change in pupil diameter between 6–8 mm within the
L∗≧ 25 range for subsequent calculations. This range was chosen based on the experiment by Bradley et al. [
2], which determined that the pupil diameter tends to decrease with age. The pupil diameter of observers in their 20s in this experiment ranged from 5.7 mm to 8.8 mm. The
L∗ range was referenced from the range of the binary classification line obtained using SVM.
In Eq. (
4), we substitute the pupil diameter
r, which linearly changes between 6 mm and 8 mm for
L∗≧ 25, the surrounding luminance
L′ = 0.11 cd/m
2, when displaying (
R,
G,
B) = (0,0,0), and the surrounding radius
A = 20 mm when considering a circle cut out of a blackout curtain as the extended light source projected on the retina. Figure
11 shows the functions obtained by further substituting the luminance and radius of reference points 1, 2, and 3 into the resulting equation, which is plotted in black in Fig.
8. Fig.
11(a), (b), and (c) correspond to reference point 1, 2, and 3, respectively.
The similarity between the binary classification line, which indicates the threshold for points perceived to be closer or farther than the reference points, and the function showing the relationship between L∗ and the viewing angle when the retinal illuminance is equal to that of the reference point, suggests that the level of retinal illuminance may determine the depth perception of small-field light sources. Since the values of retinal illuminance, area of the image projected onto the retina, including the surroundings of small-field light source and pupil diameter are based on assumptions, there is a need for verification through accurate measurements. Assuming that the depth and the magnitude of retinal illuminance determine the perception of small-field light sources, and the observers’ pupil sizes were measured, the identical retinal illuminance function obtained can accurately determine depth.